Instructions:
Air and steam mixture sterilizer adopts air and steam mixed as sterilizing medium to do sterilization, air and steam are fully mixed through fan, and circulate in a sterilizer according to a specific route. At the same time, good heat transfer is provided to ensure the uniformity of temperature, and the drying effect is better. By controlling the relationship between pressure and temperature, the phenomenon of sterilizing products such as foaming, deforming and plugs are avoided.
Structure and Features:
1. The main body is composed of inner layer and outer layer. The inner chamber is made of high quality acid-proof 304 stainless steel. Automatic welding.
2. It adopts motor-driven sealing door. Operation easily and safely. The sealing strips adopt excellent heat-resistant silicon rubber with long service life.
3. The control system adopts computer panel, PLC to ensure running steadily and inspecting temperature correctly.
4. Complete functions of date record, printing and storing. Temperature, time and pressure graph may be printed.
5. Two ways recorders.
6. Pipeline system adopts the most excellent accessories in the trade, such as GEMU air-operated valve from Germany and NASH vacuum pump and so on.
7. Mechanical seal adopts the German brand to ensure the operation is stable and safe.
Steam pressure: 0.3~0.5 Mpa
Design pressure: 0.25 Mpa
Water pressure: 0.15~0.3 Mpa
Working pressure: 0.23 Mpa
Compressed air pressure: 0.5~0.7 Mpa
Working temperature: 134 ℃